The target fluid and pump conditions for the NeoVAD Masters Dissertation are detailed below.
| Patient Weight (kg) | Flow Rate (L/min) | Pressure Head (mmHg) | Pump Speed (rpm) |
|---|---|---|---|
| 5 | 0.5 | 52 | 12250 |
| 10 | 1 | 60 | 14250 |
| 20 | 1.8 | 62 | 17750 |
The pump performance curves with the key operating conditions (24)
The curves were obtained through numerous hydraulic tests which measured the pressure and flow generated by the Impeller designs against a range of systemic vascular resistances, which were simulated using a pinch valve.